Comparing MA300 EMG System Features
| This page helps you compare the differences and similarities between each MA300 system. You can print this page out and use it to compare a Motion Lab Systems EMG system to any other manufacturer's system. We are confident that our EMG systems will beat any competing EMG system when you compare the price and specifications of one system against another. |
| Features | MA300-XII | MA300-XVI | MA300-6 | MA300-10 | MA300-16 | Comparison |
| Call for a current price | $ | $ | $ | $ | $ | $ |
| Total number of signal inputs | 12 | 16 | 18 | 22 | 28 | |
| No. of EMG channels | 8 | 16 | 6 | 10 | 16 | |
| Dedicated event switch channels | 0 | 8 | ||||
| No. of dedicated AUX channels | 4 | 0 | 4 | |||
| Number of EMG gain settings | 10 individual selections per channel | |||||
| Overall EMG Amplification | 350 to 18k ±5% | |||||
| EMG signal delay | < 2ms | < 1.2 ms | ||||
| Maximum EMG Bandwidth | 10-1000 Hz -3dB | 10-2000 Hz -3dB | ||||
| Adjacent channel cross talk | > -50 dB | |||||
| AUX Bandwidth | DC-120Hz | n/a | DC-120Hz | |||
| EMG signal bandwidth roll off | 10th Order Bessel linear phase | |||||
| Variable Low Pass EMG filter | 1,000 Hz | variable 350Hz to 2,000Hz | ||||
| Number of LP filter settings | none | 8 | ||||
| Variable High Pass EMG filter | optional | |||||
| Maximum EMG Frequency Range | 10 to 1000 Hz | 10 to 2000 Hz | ||||
| Minimum EMG Frequency Range | 10 to 1000 Hz | 20 to 350 Hz | ||||
| Group Delay (input to output) | less than 2ms | less than 2ms | ||||
| Suitable for fine-wire EMG | YES | |||||
| Withstands fine-wire stimulation | YES | |||||
| Surface contact electrodes | YES | |||||
| Gel "snap" EMG electrodes | YES | |||||
| EMG preamplifier gain | x20 | |||||
| EMG input impedance | > 100 M ohm | |||||
| Overall Noise (RMS, R.T.I) | < 1.2 uV | |||||
| C.M.R.R. | 120 dB @ 40Hz | |||||
| Electrostatic protection | YES | |||||
| Leakage Current | < 100 uA | |||||
| Subject Isolation | > 3000 Volts RMS (AC) | |||||
| Maximum output voltage (V) | Industry Standard ± 5.0V | |||||
| Mechanical Specifications - Main amplifier | ||||||
| Dimensions | 305x305x64mm | |||||
| Enclosure Material | Fully Shielded Solid Steel Enclosure | |||||
| Output Connector | Industry Standard DB-25 connector | |||||
| Output cable supplied with system | YES (BNC or pin connections) | |||||
| Mechanical Specifications - Subject unit | ||||||
| Dimensions | 137x105x45mm | |||||
| Mass | 390 g | 400 g | 420 g | 440 g | 460 g | |
| Enclosure Material | 6061-T4 Aluminum | |||||
| Connector Material | Plastic | Metal | ||||
| Connecting Cable (subject) | ||||||
| Digital signal transmission | YES | |||||
| Length | 18m (60 feet) to 36m (100 feet) | |||||
| Diameter | One single coaxial cable 2.66 mm diameter | |||||
| Usability Features | ||||||
| Universal 100/240V AC Power supply | YES | |||||
| Internal calibration source | YES | |||||
| EMG overload LEDs | YES | |||||
| Signal OK / CRC error detection | YES | |||||
| Biomechanics standard C3D files | YES | |||||
| Regulatory Approvals | ||||||
| UL approval | YES | |||||
| IEC 601-1 CE approval | YES | |||||
| Warranty | ||||||
| Money Back Guarantee | YES | |||||
| Warranty is Transferable | YES | |||||
| Standard Warranty Period | 2 years | |||||